PUMP IN STYLE ADVANCED
Report
- Report Number
- 1419937-2012-00736
- Event Type
- Injury
- Date Received
- November 16, 2012
- Date of Event
- October 20, 2012
- Report Date
- October 20, 2012
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE CUSTOMER WAS SENT A REPLACEMENT PUMP IN STYLE. IN FOLLOW UP WITH THE CUSTOMER ON (B)(4) 2012, THE CUSTOMER INDICATED THAT HER MILK EXPRESSION HAD DECREASED SUDDENLY AND SHE BELIEVED THAT THE PUMP IN STYLE DEVICE THAT SHE WAS USING HAD LOWER SUCTION. SHE DID CONSULT HER HEALTHCARE PROVIDER WAS DIAGNOSED WITH MASTITIS AND WAS PRESCRIBED CEPHALEXIN, AN ANTIBIOTIC. SHE ALSO STATED THAT THE MASTITIS WAS MUCH IMPROVED AFTER TAKING THE ANTIBIOTIC. THE CUSTOMER WAS ALSO CONTACTED BY A MEDELA CLINICIAN ON (B)(4) 2012. THE CUSTOMER REPORTED TO THE CLINICIAN THAT THE MASTITIS WAS CLEARING UP AND THAT THE REPLACEMENT PUMP SHE RECEIVED FROM MEDELA WAS WORKING WELL. "MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION, WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OF MASTITIS." ["BREASTFEEDING AND HUMAN LACTATION" (RIORDAN AND WAMBACH, 4TH EDITION, PAGE 294)]. THE BREAST PUMP WAS RETURNED BY THE CUSTOMER FOR TESTING/ANALYSIS. BASED ON THE FACT THAT THE PUMP TESTED TO SPECIFICATION, IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE MASTITIS. WE WILL MONITOR COMPLAINTS FOR SIMILAR ISSUES. SHOULD ADDITIONAL INFORMATION BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW UP REPORT WILL BE FILED. EVALUATION SUMMARY: THE PUMP IN STYLE DEVICE WAS RETURNED FOR TESTING WITH THE KIT, BUT NO MEMBRANES. THE PUMP WAS TESTED, NO PROBLEMS WERE FOUND. THE PUMP WAS FOUND TO BE WITHIN VACUUM SPECIFICATIONS.
THE CUSTOMER REPORTED THAT SHE WAS EXPERIENCING LOW MILK EXPRESSION AND THAT SHE WAS DIAGNOSED WITH MASTITIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP IN STYLE ADVANCED | HGX | MEDELA, INC. | 57063 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |