COULTER® LH 750 ANALYTICAL STATION
Report
- Report Number
- 1061932-2012-02690
- Event Type
- Malfunction
- Date Received
- November 25, 2012
- Date of Event
- October 31, 2012
- Report Date
- October 31, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE HAD OBSERVED THERE WAS AN ISOTON POOLED AROUND THE DIFFERENTIAL (DIFF) MODULE. THE SOURCE OF THE LEAK WAS WORN PINCH VALVE (VL) 46B. THE VL46 SUPPLIES PRESSURIZED DILUENT TO RINSE THE UPPER CHAMBER IN THE FLOW CELL. THE FSE CLEANED AND DRIED THE LIGHT SCATTER PREAMP WHICH HAD GOTTEN LEAKED ON FROM THE WORN TUBING. ONCE DRIED, THE PREAMP WORKED PROPERLY. THE FSE ALSO CLEANED THE REST OF THE FLUID LEAK FROM AROUND THE DIFF TOWER. NO FURTHER EVIDENCE OF LEAKING WAS OBSERVED, INSTRUMENT RAN WITHOUT INCIDENT. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BEC) AND REPORTED THAT THE COULTER LH 750 ANALYTICAL STATION WAS GENERATING LIGHT SCATTER (LS) OFFSET ERRORS AND R FLAGS FOR THE DIFFERENTIALS WHILE THE CUSTOMER WAS RUNNING QC. THE CUSTOMER POWERED THE UNIT OFF/ON AND OBSERVED A LEAK. THE VOLUME OF THE LEAK WAS APPROXIMATELY 1 CUP AND NOT CONTAINED WITHIN THE INSTRUMENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LAB'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, EYE PROTECTION, AND A LAB COAT AT THE TIME OF THE EVENT. NO ONE CAME IN CONTACT WITH THE FLUID. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. FLUIDS ASSOCIATED WITH THIS EVENT WERE DILUENT AND CLENZ. NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 750 ANALYTICAL STATION | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |