ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00846
- Event Type
- Malfunction
- Date Received
- November 25, 2012
- Report Date
- November 1, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT206 IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: ONLY THE COMPLAINT RT206 DRYLINE WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR INSPECTION. THE TUBING WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A HOLE IN THE DRYLINE, ABOUT 16CM AWAY FROM THE CONNECTOR. A LOT CHECK WAS NOT PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: IT WAS IDENTIFIED THAT THE DAMAGE TO THE DRYLINE TUBE WAS CAUSED BY A FAULTY CORRUGATOR BLOCK THAT WAS USED DURING THE DRYLINE MANUFACTURING PROCESS. THE FAULTY BLOCK WAS REPLACED IN (B)(6) 2012. WE HAVE NOT RECEIVED ANY COMPLAINTS OF THIS NATURE FOR PRODUCT MANUFACTURED AFTER (B)(6) 2012. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY 10 TO 15 MINUTES AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT105 ADULT INSPIRATORY-HEATED BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).
A DISTRIBUTOR IN (B)(6) REPORTED THAT THE DRYLINE OF AN RT206 ADULT BREATHING CIRCUIT HAD A HOLE IN IT. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |