FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2843804
·
Received October 26, 2012
Report
- Report Number
- 3008642652-2012-02748
- Event Type
- Malfunction
- Date Received
- October 26, 2012
- Date of Event
- September 23, 2012
- Report Date
- October 16, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) IS CURRENTLY UNDERWAY. THE REPORTED PROBLEM (WILL NOT POWER UP) HAS BEEN CONFIRMED. UPON EVAL THE MONITOR WOULD NOT POWER ON. UPON INVESTIGATION, THERE WAS A SEGMENTATION FAULT WHILE PERFORMING THE FLASH MEMORY TEST. THE CAUSE OF THE PROBLEM WAS ISOLATED TO FLASH MEMORY COMPONENTS U102 AND U105 ON COMPUTER ANALOG (CA) BOARD SN (B)(4). A ROOT CAUSE ANALYSIS IS CURRENTLY UNDERWAY. A SUPPLEMENTAL REPORT WILL BE SENT UPON COMPLETION OF EVAL. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULD NOT POWER UP PROPERLY. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |