6 SHOOTER SAEED MULTI-BAND LIGATOR
Report
- Report Number
- 1037905-2012-00593
- Event Type
- Malfunction
- Date Received
- October 25, 2012
- Date of Event
- September 24, 2012
- Report Date
- September 25, 2012
- Manufacturer
- WILSON-COOK MEDICAL INC.
- Product Code
- MND
- PMA / PMN Number
- K944220
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
OF THE THREE (3) DEVICES, ONLY ONE(1) DEVICE WAS MADE AVAILABLE FOR EVAL; 1037905-2012-00591: RETURNED ON (B)(4) 2012, 1037905-2012-00592: NOT AVAILABLE FOR EVAL, 1037905-2012-00593. INVESTIGATION EVAL: A PRODUCT EVAL WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVAL. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: THE INSTRUCTIONS FOR USE DIRECT THE USER TO ATTACH THE TRIGGER CORD TO THE HOOK ON THE END OF THE LOADING CATHETER, LEAVING APPROX 2CM OF TRIGGER CORD BETWEEN THE KNOT AND THE HOOK. IF THE KNOT AND THE HOOK ARE PLACED CLOSER THAN 2CM, THIS CAN CREATE RESISTANCE WHEN WITHDRAWING THE LOADING CATHETER THROUGH THE LIGATOR HANDLE AND/OR ACCESSORY CHANNEL OF THE ENDOSCOPE. IF ADD¿L PRESSURE IS APPLIED TO THE LOADING CATHETER WHEN RESISTANCE IS ENCOUNTERED, THIS COULD CONTRIBUTE TO SEPARATION OF THE BLUE HOOK FROM THE LOADING CATHETER. PRIOR TO DISTRIBUTION, ALL 6 SHOOTER SAEED MULTI-BAND LIGATORS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN IMPLANTED TO REDUCE OCCURRENCES OF BLUE HOOK DETACHMENT. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
IN PREPARATION FOR AN ENDOSCOPY PROCEDURE, A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR WAS USED. THE USER WANTED TO MAKE BANDING WITH THE MBL, BUT THE BLUE HACK [HOOK] PEELED OFF FROM THE [LOADING] CATHETER FROM THE SYSTEM. THE CUSTOMER USED A NEW COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR FROM ANOTHER LOT NUMBER, THIS WAS OK. THE SAME OBSERVATION WAS MADE WITH THREE (3) LIGATORS. SEE MDRS 1037905-2012-00591, 1037905-2012-00592 AND 1037905-2012-00593. THIS OCCURRED DURING THE LOADING PROCESS; THE LOADING CATHETER WAS NOT LOCATED INSIDE THE PT¿S BODY. THE PT DID NOT REQUIRE ANY ADD¿L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6 SHOOTER SAEED MULTI-BAND LIGATOR | MND, LIGATOR, ESOPHAGEAL | MND | WILSON-COOK MEDICAL INC. | W3145675 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS (B)(4) ENDOSCOPE |