FDA Adverse Event Malfunction Summary report: N

OLYMPUS SINGLE USE MECHANICAL LITHOTRIPTOR V

MDR report key: 2843755 · Received October 25, 2012

Report

Report Number
8010047-2012-00407
Event Type
Malfunction
Date Received
October 25, 2012
Date of Event
September 27, 2012
Report Date
September 27, 2012
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
FEO
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

OLYMPUS FOLLOWED-UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFO REGARDING THIS REPORT. THE USER FACILITY REPORTED THAT THE PT WAS AN IN-PATIENT AND A GALL BLADDER SURGERY HAD BEEN SCHEDULED TO BE PERFORMED THE FOLLOWING DAY. THE USER FACILITY REPORTED THAT THE PT'S COMMON BILE DUCT WAS FILLED WITH STONES, AND 1/3 OF THE STONES WERE REMOVED WHILE THE REMAINDER STONES WERE "STACK ON TOP OF ONE ANOTHER". THE USER FACILITY REPORTED THAT THE PT WAS VISITED ON (B)(6) 2012 AND WAS FOUND DOING WELL. THE DEVICE REFERENCED IN THIS REPORT WAS NOT RETURNED TO OLYMPUS FOR EVAL. THE EXACT CAUSE OF THE USER'S EXPERIENCE COULD NOT BE CONCLUSIVELY DETERMINE. HOWEVER, THE SIZE, HARDNESS AND EXCESSIVE QUANTITY OF STONES COULD NOT BE RULED OUT AS A CONTRIBUTORY FACTOR. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE REFERENCED DEVICE WAS WORKING FINE BUT POSSIBLY ENCOUNTERED A LARGE AND HARD STONE WHICH CAUSED THE SHEATH, WIRE BASKET, AND THE ENDOSCOPE TO BE STUCK INSIDE THE PT. THE USERS EVENTUALLY MANAGED TO CUT AND REMOVE THE SHEATH AND THE ENDOSCOPE WITH THE BASKET REMAINING. THE PT REPORTEDLY HAD TO UNDERGO AN OPEN SURGERY FOR BASKET REMOVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPUS SINGLE USE MECHANICAL LITHOTRIPTOR V MECHANICAL LITHOTRIPTOR FEO OLYMPUS MEDICAL SYSTEMS CORPORATION BML-V442QR-26 1ZK

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention OLYMPUS TJF-180V ENDOSCOPE WITH UNK SERIAL NUMBER