FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2843732 · Received October 26, 2012

Report

Report Number
1314492-2012-00379
Event Type
Malfunction
Date Received
October 26, 2012
Report Date
September 28, 2012
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SIGMA DEVICE EVALUATION FOUND THE #2, #4, #5, #6, #7, #8, #9, AND (.) KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #2 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 12 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AD WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. SIGMA DEVICE EVALUATION CONFIRMED THAT THE COMPLAINT "DOOR IS HARD TO CLOSE WITH A LOADED IV SET" WAS CAUSED BY CUSTOMER DAMAGE TO THE DOOR HOOKS. THIS REPORT IS BEING SUBMITTED DUE TO THE POTENTIAL FOR PATIENT INJURY THAT COULD RESULT FROM THE DOUBLE OUTPUT OF THE KEYPAD. THE DATE OF EVENT IS UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP STATED "KEYPAD 3RD + 4TH ROW". IT WAS ALSO REPORTED THAT THE DOOR IS HARD TO CLOSE WITH A LOADED IV SET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1