FDA Adverse Event Malfunction Summary report: N

OHMEDA QUICK CONNECT

MDR report key: 2843524 · Received September 21, 2012

Report

Report Number
2843524
Event Type
Malfunction
Date Received
September 21, 2012
Date of Event
July 19, 2012
Report Date
September 21, 2012
Manufacturer
CARDINAL HEALTH
Product Code
BXY
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

PATIENT WAS CONNECTED TO MEDICAL AIR AND NOT OXYGEN. WE USE PISS (PIN INDEX SAFETY SYSTEM) FLOWMETERS IN THE NEWER AREAS OF THE HOSPITAL AND DISS(DIAMETER INDEX SAFETY SYSTEM) FLOWMETERS IN THE OLDER WINGS. EITHER WAY YOU MAY NOT PLUG AN OXYGEN FLOWMETER INTO AN AIR OUTLET OR VICE VERSA. THE NIPPLE ADAPTER THAT IS USED TO ATTACH A DELIVERY DEVICE (E.G. MASK, NASAL CANNULA) IS THE SAME SIZE. WE HAVE GREEN/YELLOW COLOR CODED NIPPLES WHICH ARE INTERCHANGEABLE. THE PROBLEM ARISES WHEN STAFF PLACES THE INCORRECT NIPPLE ONTO THE FLOWMETER AND ASSUMES THAT THE NIPPLE COLOR DENOTES THE DELIVERED GAS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OHMEDA QUICK CONNECT FLOWMETER, CALIBRATION, GAS, ADAPTER BXY CARDINAL HEALTH * *
2 OHMEDA QUICK CONNECT AIR FLOWMETER, CALIBRATION, GAS, CONNECTOR BXY CARDINAL HEALTH * *
3 PRECISION MEDICAL 02 FLOW METER FLOWMETER, CALIBRATION, GAS, ADAPTER BXY CARDINAL HEALTH * *
4 * FLOWMETER, CALIBRATION, GAS, ADAPTER BXY CAREFUSION * *