FDA Adverse Event
Malfunction
Summary report: N
OHMEDA QUICK CONNECT
MDR report key: 2843524
·
Received September 21, 2012
Report
- Report Number
- 2843524
- Event Type
- Malfunction
- Date Received
- September 21, 2012
- Date of Event
- July 19, 2012
- Report Date
- September 21, 2012
- Manufacturer
- CARDINAL HEALTH
- Product Code
- BXY
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
Narratives
Description of Event or Problem · 1
PATIENT WAS CONNECTED TO MEDICAL AIR AND NOT OXYGEN. WE USE PISS (PIN INDEX SAFETY SYSTEM) FLOWMETERS IN THE NEWER AREAS OF THE HOSPITAL AND DISS(DIAMETER INDEX SAFETY SYSTEM) FLOWMETERS IN THE OLDER WINGS. EITHER WAY YOU MAY NOT PLUG AN OXYGEN FLOWMETER INTO AN AIR OUTLET OR VICE VERSA. THE NIPPLE ADAPTER THAT IS USED TO ATTACH A DELIVERY DEVICE (E.G. MASK, NASAL CANNULA) IS THE SAME SIZE. WE HAVE GREEN/YELLOW COLOR CODED NIPPLES WHICH ARE INTERCHANGEABLE. THE PROBLEM ARISES WHEN STAFF PLACES THE INCORRECT NIPPLE ONTO THE FLOWMETER AND ASSUMES THAT THE NIPPLE COLOR DENOTES THE DELIVERED GAS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OHMEDA QUICK CONNECT | FLOWMETER, CALIBRATION, GAS, ADAPTER | BXY | CARDINAL HEALTH | * | * | |
| 2 | OHMEDA QUICK CONNECT AIR | FLOWMETER, CALIBRATION, GAS, CONNECTOR | BXY | CARDINAL HEALTH | * | * | |
| 3 | PRECISION MEDICAL 02 FLOW METER | FLOWMETER, CALIBRATION, GAS, ADAPTER | BXY | CARDINAL HEALTH | * | * | |
| 4 | * | FLOWMETER, CALIBRATION, GAS, ADAPTER | BXY | CAREFUSION | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |