Report
- Report Number
- 3008382007-2012-06613
- Event Type
- Malfunction
- Date Received
- November 23, 2012
- Report Date
- November 19, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K082590
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2012, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING A ONETOUCH PING METER WAS READING INACCURATELY HIGH RESULTS COMPARED TO A CALIBRATED LAB DRAW. THE PATIENT REPORTED OBTAINING A BLOOD GLUCOSE VALUE OF "214MG/DL" ON THE LFS METER AND "163MG/DL" ON THE LAB DRAW. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE BETWEEN THE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20MG/DL OR <=20% OBTAINED WITHIN 30 MINUTES OF EACH OTHER. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY DUE TO THE REPORTED ISSUE. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THE SUBJECT METER DID NOT CAUSE OR CONTRIBUTE TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE READINGS, SYMPTOMS OR TREATMENT SUGGESTIVE THAT A SERIOUS INJURY OCCURRED. HOWEVER THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |