FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 2843367 · Received November 23, 2012

Report

Report Number
3008382007-2012-06613
Event Type
Malfunction
Date Received
November 23, 2012
Report Date
November 19, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING A ONETOUCH PING METER WAS READING INACCURATELY HIGH RESULTS COMPARED TO A CALIBRATED LAB DRAW. THE PATIENT REPORTED OBTAINING A BLOOD GLUCOSE VALUE OF "214MG/DL" ON THE LFS METER AND "163MG/DL" ON THE LAB DRAW. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE BETWEEN THE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20MG/DL OR <=20% OBTAINED WITHIN 30 MINUTES OF EACH OTHER. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY DUE TO THE REPORTED ISSUE. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THE SUBJECT METER DID NOT CAUSE OR CONTRIBUTE TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE READINGS, SYMPTOMS OR TREATMENT SUGGESTIVE THAT A SERIOUS INJURY OCCURRED. HOWEVER THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1 63 YR