FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2843225 · Received November 21, 2012

Report

Report Number
2955842-2012-01060
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 22, 2012
Report Date
October 26, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE FOUND TO THE INSTRUMENT'S PITCH CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE WIRES ON THE MEGASUTURECUT NEEDLE DRIVER INSTRUMENT FRAYED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120723 728

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES.