FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 2843058 · Received November 21, 2012

Report

Report Number
9611451-2012-00841
Event Type
Malfunction
Date Received
November 21, 2012
Report Date
November 2, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT380 ADULT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE INSPIRATORY AND EXPIRATORY HEATER WIRES OF THE RETURNED RT380 ADULT BREATHING CIRCUIT WERE VISUALLY INSPECTED FOR DAMAGED PINS AND TESTED TO SEE IF IT COULD BE CONNECTED TO A HEATER WIRE ADAPTOR. RESULTS: A HEATER WIRE ADAPTOR COULD NOT BE FULLY CONNECTED TO THE HEATER WIRE SOCKET IN THE INSPIRATORY TUBE. VISUAL INSPECTION REVEALED THAT THE INSPIRATORY HEATER WIRE PINS WERE BENT. THIS PREVENTS THE ADAPTOR FROM CONNECTING TO THE HEATER WIRE SOCKET. NO FAULT WAS FOUND WITH THE EXPIRATORY HEATER WIRES. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120918. CONCLUSION: IT IS POSSIBLE FOR THE USER TO DAMAGE THE HEATER WIRE PINS DURING USE, FOR EXAMPLE, IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR BENT OR SPLIT PINS REPORTED TO US BY HEALTH CARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE DAMAGED DURING PRODUCTION OR BY THE END USER. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT AN RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT COULD NOT BE CONNECTED TO THE HEATER WIRE ADAPTOR. THIS WAS NOTICED BEFORE PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT380 120918

Patients

Seq Age Sex Outcome Treatment
1