FDA Adverse Event Malfunction Summary report: N

HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE

MDR report key: 2842955 · Received November 21, 2012

Report

Report Number
2024168-2012-07388
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 30, 2012
Report Date
October 30, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
DQX
PMA / PMN Number
K103101
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. (B)(4) - EXCESSIVE FORCE. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE COIL DAMAGE WAS ABLE TO BE CONFIRMED HOWEVER THE RESISTANCE WITH THE SDS WAS UNABLE TO BE CONFIRMED. BASED ON A VISUAL, DIMENSIONAL, AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WS REPORTED THAT DURING ABBOTT VASCULAR INTERNAL BENCHTOP TESTING WITH A SYNTHETIC ARTERIAL MODEL, THE BALANCE MIDDLEWEIGHT (BMW) ELITE GUIDE WIRE WAS USED WITH A 3.25 X 38 MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS). DURING ATTEMPTED REMOVAL THE BMW ELITE BECAME STUCK IN THE CATHETER AND COULD NOT BE REMOVED. IT REQUIRED 'EXCESSIVE FORCE' TO REMOVE THE GUIDE WIRE FROM THE SDS. THE GUIDE WIRE WAS REMOVED IN ONE PIECE, AND IT WAS NOTED THAT THE COILS ON THE BMW ELITE APPEARED OFFSET. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO NOTED ISSUE WITH THE SDS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE GUIDE WIRE DQX AV-TEMECULA-CT 2060701

Patients

Seq Age Sex Outcome Treatment
1 STENT: XIENCE XPEDITION SDS