Summary report: N
FENESTRATED BIPOLAR FORCEPS INSTRUMENT
MDR report key: 2842849
·
Received November 21, 2012
Report
- Report Number
- 2955842-2012-01040
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 25, 2012
- Report Date
- October 26, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. CUSTOMER INITIALLY REPORTED EXPOSED WIRES. HOWEVER FOR CLARIFICATION, THE PRODUCT PROBLEM WAS A BROKEN PITCH CABLE. BASED ON THIS, THE COMPLAINT WAS CONFIRMED. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT HAD EXPOSED WIRES. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.