FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2842793 · Received November 21, 2012

Report

Report Number
2955842-2012-01035
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 23, 2012
Report Date
October 25, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT OF BROKEN CABLE WAS CONFIRMED. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING A BROKEN CABLE STICKING OUT AT THE DISTAL END OF THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120514 759

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES