ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2012-05296
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 22, 2012
- Report Date
- November 7, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS. SHOULD THE DEVICE BE RETURNED FOR ANALYSIS, A SUPPLEMENTAL MEDWATCH WILL BE SENT
(B)(4). THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED - EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE 'INSTRUMENT ERROR' ALERT WAS DISPLAYED; AND WHEN TESTED ON THE GEN04 GENERATOR, AN ERROR CODE 5 (INSTRUMENT ERROR) WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE TO STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 OR INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS 'REPLACE INSTRUMENT' WITH THE GEN11 LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.
(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE USING A GENERATOR, THE JAW OF THE DEVICE BROKE. THE BROKEN PART HAS BEEN RETRIEVED BY THE SURGEON. THE CASE WAS COMPLETED BY USING ANOTHER DEVICE. CUSTOMER CANNOT PROVIDE ANY MORE DETAILS ABOUT THE CIRCUMSTANCES IN WHICH THE ISSUE OCCURRED. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |