FDA Adverse Event
Malfunction
Summary report: N
AFFINITY IV BED
MDR report key: 2842697
·
Received November 13, 2012
Report
- Report Number
- 1824206-2012-07397
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Date of Event
- October 23, 2012
- Report Date
- October 23, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE BOLT BROKEN FOR THE TRENDELENBURG LINKAGE. THE TECHNICIAN REPLACED THE BOLT FOR THE TRENDELENBURG LINKAGE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THAT THE BED WILL NOT GO INTO TRENDELENBURG. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY IV BED | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |