FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2842668 · Received November 21, 2012

Report

Report Number
1416980-2012-05752
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 9, 2012
Report Date
November 10, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4).THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) IS CONFIRMED BECAUSE THE PATIENT REPORTED HAVING A CLAMP OPEN ON AN UNUSED SUPPLY LINE DURING THERAPY, WHICH IS USE ERROR THAT CAN CAUSE SYSTEM ERROR 2240 ALARMS. THE HOME PATIENT (HP) STATED THE FINAL LINE WAS IN THE ORGANIZER UNUSED AND THE BLUE CLAMP WAS OPEN. A LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240 (AIR IN SET) DURING FILL 6 OF 7 ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) STATED THE FINAL LINE WAS IN THE ORGANIZER UNUSED AND THE BLUE CLAMP WAS OPEN AND THE HP KNEW IT SHOULD HAVE BEEN CLOSED. THE HP WOULD NOT CONTINUE THERAPY THAT NIGHT. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP TO END THERAPY AND ADVISED THE HP TO DISPOSE OF THE CURRENT SUPPLIES USED. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOMECHOICE