FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION

MDR report key: 2842651 · Received November 21, 2012

Report

Report Number
2032227-2012-07806
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 6, 2012
Report Date
November 6, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

UNRESPONSIVE BUTTONS DUE TO MOISTURE DAMAGE ON KEYPAD TRACE. DISPLAY FUNCTIONS PROPERLY WITH TESTING CASE. NO FROZEN DISPLAY NOTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY. CUSTOMER STATED THAT THE TIME ON THE INSULIN PUMP IS NOT ADVANCING AND THERE IS NO RESPONSE FROM ANY BUTTON. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION OYC OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 37 YR