FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2842617 · Received November 21, 2012

Report

Report Number
2531779-2012-13842
Event Type
Malfunction
Date Received
November 21, 2012
Report Date
October 25, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED MULTIPLE LOSS OF PRIME WARNINGS WITH BOTH ZERO AND NON ZERO LOW FORCE. THE PUMP WAS EXERCISED FOR 24 HOURS ON A ONE UNIT BASAL RATE WITH NO LOSS OF PRIMES DUPLICATED. THE FORCE CALIBRATION TEST FAILED. THERE WAS MOISTURE CONTAMINATION FOUND IN FORCE SENSOR HOUSING AND FORCE SENSOR SHIM. THE PUMP WAS OPENED AND CONTAMINATION WAS FOUND ON THE FORCE SENSOR PINS AND IN THE PUMP. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE FORCE CALIBRATION TEST FAILED INVESTIGATION. THERE WAS CONTAMINATION FOUND ON THE FORCE SENSOR PINS AND IN THE PUMP. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 22 YR