FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 2842615 · Received November 21, 2012

Report

Report Number
2024168-2012-07378
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 25, 2012
Report Date
October 31, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FAILURE TO FOLLOW STEPS/INSTRUCTIONS. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE STENT DISLODGMENT WAS CONFIRMED. BASED ON VISUAL, FUNCTIONAL, DIMENSIONAL AND SCANNING ELECTRON MICROSCOPY (SEM) ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. IT SHOULD BE NOTED THAT THE XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE (IFU) STATES: CAREFULLY REMOVE THE DELIVERY SYSTEM FROM ITS PROTECTIVE TUBING FOR PREPARATION OF THE DELIVERY SYSTEM. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RETURNED FOR INVESTIGATION. THE INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREP, WHEN REMOVING THE PROTECTIVE SHEATH, THE STENT IMPLANT DISLODGED. THERE WAS NO RESISTANCE REMOVING THE SHEATH. ANOTHER XIENCE PRIME WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2030741

Patients

Seq Age Sex Outcome Treatment
1