LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Report
- Report Number
- 3015876-2012-00851
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 22, 2012
- Report Date
- October 24, 2012
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND WAS UNABLE TO VERIFY OR REPLICATE THE REPORTED PROBLEM. ONE OF THE BATTERY POSTS WAS LOOSE BUT PROPER DEVICE OPERATION OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. A CONCLUSIVE CAUSE OF THE REPORTED EVENT WAS NOT DETERMINED. THE THIRD PARTY SERVICE AGENT EVALUATED THE DEVICE AND WAS UNABLE TO VERIFY OR DUPLICATE THE REPORTED PROBLEM. THE SERVICE AGENT DID OBSERVED A BATTERY PIN TO BE LOOSE. THE SERVICE AGENT TIGHTENED THE BATTERY PIN AND PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. A CONCLUSIVE CAUSE OF THE REPORTED EVENT WAS NOT DETERMINED.
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND WAS UNABLE TO VERIFY OR REPLICATE THE REPORTED PROBLEM. ONE OF THE BATTERY POSTS WAS LOOSE BUT PROPER DEVICE OPERATION OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. A CONCLUSIVE CAUSE OF THE REPORTED EVENT WAS NOT DETERMINED.
IT WAS REPORTED THAT THE DEVICE INTERMITTENTLY TURNS ITSELF OFF. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.