FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 2842352 · Received October 22, 2012

Report

Report Number
1220908-2012-02815
Event Type
Malfunction
Date Received
October 22, 2012
Report Date
October 4, 2012
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS NOT RECEIVED THE PRODUCT FOR EVAL AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA