FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 2842344 · Received November 21, 2012

Report

Report Number
3004209178-2012-90865
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 18, 2012
Report Date
November 18, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATIONS. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIR CONNECTED AND LOCKED IN PLACE PROPERLY.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN FROM THE RESERVOIR WAS LEAKING PAST THE O-RINGS. TROUBLESHOOTING WAS PERFORMED. ADVISED THE CALLER WHEN FILLING THE RESERVOIR WITH AIR SHE DOES NOT NEED TO PUSH THE PLUNGER BACK AND FORT. NO FURTHER INFORMATION WAS PROVIDED.