PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2012-90867
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 16, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED PRIME, DISPLACEMENT AND BASIC OCCLUSION TESTS. NO PRIME ALARMS OR DISPLACEMENT ANOMALY NOTED. THE INSULIN PUMP RECEIVED STUCK IN MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. MOTOR MAY HAVE HAD INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING TESTING. NO PRIME ALARMS OR DISPLACEMENT ANOMALY NOTED.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. IT WAS STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CALLER TO RUN THE DISPLACEMENT TEST AND THE TEST FAILED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |