FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 2842192 · Received October 22, 2012

Report

Report Number
1722139-2012-01067
Event Type
Malfunction
Date Received
October 22, 2012
Date of Event
October 2, 2012
Report Date
October 5, 2012
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PUMP WAS TESTED FOR ACCURACY SEVERAL TIMES, USING WATER. EACH TIME PUMP DELIVERED AMOUNT WITHIN SPECIFICATION ((B)(4)). COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

CUSTOMER STATES THAT PUMP DELIVERED 12 ML DURING TESTING. RATE, DOSE, AND FOOD ARE 500 ML/HR, 10 ML, AND WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MEDICAL DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1