DUROM ACETABULAR COMPONENT 48/42 CODE H

MDR report key: 2842115 · Received November 13, 2012

Report

Report Number
9613350-2012-01065
Event Type
Injury
Date Received
November 13, 2012
Date of Event
March 11, 2011
Report Date
October 22, 2012
Manufacturer
ZIMMER GMBH
Product Code
KWY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER DID NOT RECEIVED DEVICES OR X-RAYS FOR REVIEW. AS NOT LOT NUMBERS WERE PROVIDED FOR THE DEVICES, THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED. WITH THE INFORMATION GIVEN SO FAR, NO FURTHER INVESTIGATION IS POSSIBLE. BASED ON AN EXTENSIVE INVESTIGATION OF EVENTS REPORTED FROM SEVERAL USER FACILITIES OUTSIDE THE U.S., ZIMMER IDENTIFIED THAT THE MOST PROBABLE CAUSE FOR THE OUTCOME OBSERVED WAS A LOOSE OR UNSTABLE CUP THAT RESULTED FROM USE OF SURGICAL TECHNIQUES NOT CONSISTENT WITH THE MANUFACTURER'S RECOMMENDATIONS. AS A CORRECTIVE ACTION, A RETRAINING PROGRAM FOR USERS OUTSIDE THE U.S. WAS INITIATED IN NOVEMBER 2009 AND REPORTED TO THE NATIONAL COMPETENT AUTHORITIES AS REQUIRED. THE DUROM CUP REPORTED IN THIS CASE IS NOT MARKETED IN THE U.S. A SIMILAR CUP, COMPATIBLE WITH THE METASUL LDH FEMORAL HEAD, IS CLEARED IN THE U.S. AND A CORRECTIVE ACTION FOR THIS PRODUCT WAS REPORTED TO THE FDA IN JULY 2008 AS CORRECTION Z-2415/2426-2008. SHOULD ADDITIONAL INFORMATION INCLUDING THE FINAL INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. (B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT RECEIVED AN ACETABULAR CUP AND UNDERWENT A REVISION SURGERY DUE TO PAIN. MOREOVER BURSITIS TROCHANTERICA DUE TO METAL WEAR WAS OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUROM ACETABULAR COMPONENT 48/42 CODE H DUROM ACETABULAR COMPONENT AND METASUL KWY ZIMMER GMBH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization