FDA Adverse Event Injury Summary report: N

PENTA

MDR report key: 2842057 · Received November 12, 2012

Report

Report Number
1627487-2012-06733
Event Type
Injury
Date Received
November 12, 2012
Date of Event
October 17, 2012
Report Date
October 17, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT IS NO LONGER RECEIVING ADEQUATE COVERAGE. IT WAS ALSO REPORTED THE PATIENT IS EXPERIENCING RIB STIMULATION. REPROGRAMMING ATTEMPTS WERE UNSUCCESSFUL. X-RAYS REVEALED LEAD MIGRATION. THE PATIENT WILL UNDERGO SURGICAL INTERVENTION TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PENTA SCS LEAD LGW ST. JUDE MEDICAL - NEUROMODULATION 3228 3547032

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention SCS ANCHORS: MODEL 1194(2)| IMPLANT:| IMPLANT:| SCS IPG: MODEL 3716