FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2841993 · Received November 21, 2012

Report

Report Number
1416980-2012-05722
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 31, 2012
Report Date
November 1, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K062457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 40 ML OF FLUID IN THE BLADDER. A VISUAL EXAMINATION CONFIRMED THE REPORTED CONDITION, OF THE RESERVOIR APPEARING THIN IN CERTAIN AREAS. HOWEVER, THE ROOT CAUSE COULD NOT BE DETERMINED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE SAMPLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF AN INFUSOR THAT APPEARED TO HAVE DAMAGE ON THE RESERVOIR OF THE DEVICE. THE REPORTED PROBLEM WAS NOTICED DURING FILLING. THE DEVICE WAS BEING FILLED WITH DESFERRIOXAMINE AND WATER. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12F020

Patients

Seq Age Sex Outcome Treatment
1