INFUSOR
Report
- Report Number
- 1416980-2012-05720
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 9, 2012
- Report Date
- November 1, 2012
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K071222
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A RUPTURED RESERVOIR WAS CONFIRMED. VISUAL EXAMINATION OF THE UNIT DETERMINED THAT THE BLADDER RUPTURED IN A NON-FOOTED POSITION. THE ROOT CAUSE WAS DETERMINED TO BE A MANUFACTURING DEFECT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD A RESERVOIR RUPTURE DURING FILLING. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THIS CONDITION INTERRUPTED THERAPY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 12B070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |