FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2841959 · Received November 21, 2012

Report

Report Number
1416980-2012-05720
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 9, 2012
Report Date
November 1, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A RUPTURED RESERVOIR WAS CONFIRMED. VISUAL EXAMINATION OF THE UNIT DETERMINED THAT THE BLADDER RUPTURED IN A NON-FOOTED POSITION. THE ROOT CAUSE WAS DETERMINED TO BE A MANUFACTURING DEFECT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD A RESERVOIR RUPTURE DURING FILLING. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THIS CONDITION INTERRUPTED THERAPY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12B070

Patients

Seq Age Sex Outcome Treatment
1