FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY HOOK INSTRUMENT

MDR report key: 2841864 · Received November 21, 2012

Report

Report Number
2955842-2012-01021
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 26, 2012
Report Date
October 26, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE OF FRAYED WIRES. INSTRUMENT WAS FOUND WITH BANANA PLUG BENT AT THE BACK END OF THE HOUSING. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO EXCESS FORCE APPLIED NORMAL TO THE BANANA PLUG. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, FRAYED WIRES WERE IDENTIFIED DURING SET UP ON THE PERMANENT CAUTERY HOOK INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERMANENT CAUTERY HOOK INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420183-05 M10101215 472

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES