FDA Adverse Event Malfunction Summary report: N

DRILL STOP

MDR report key: 2841272 · Received November 21, 2012

Report

Report Number
2530088-2012-01010
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 28, 2012
Report Date
October 28, 2012
Manufacturer
SYNTHES BRANDYWINE
Product Code
HTW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A PRODUCT DEVELOPMENT EVALUATION WAS CONDUCTED. THE RETURNED DRILL STOP SHOWS SIGNS OF WEAR OVER THE ENTIRE PART. THE ENTIRE PERIMETER OF THE INNER EDGE OF THE PLUNGER MECHANISM HAS SEVERAL WEAR MARKS, SMALL DINGS AND GOUGES ON IT. A SMALL PIECE OF CLEAR TAPE EXISTS ON THE MAJOR O.D. WHICH IS NOT A RESULT OF THE MANUFACTURING PROCESS.THE RETURNED DRILL STOP WAS MANUFACTURED IN AUGUST 2004 AND IS OVER 8 YEARS OLD. THE DEVICE SHOWS SIGNS OF WEAR OVER THE ENTIRE PART. THE DESIGN WAS REVIEWED, IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. THE DRILL STOP IS MADE FROM 17-4PH MATERIAL WHICH IS A STANDARD MATERIAL USED TO MAKE SYNTHES INSTRUMENTS. THE EXTENSIVE WEAR OF THE PLUNGER MECHANISM ON THE RETURNED DRILL STOP AND THE AGE OF THE DEVICE SUGGEST IT WAS USED BEYOND ITS USEFUL LIFE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF THE DHR INDICATES THERE WERE NO ANOMALIES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THIS COMPLAINT.

Description of Event or Problem · 1

DURING A TFN HIP NAILING PROCEDURE, THE SURGEON WAS REAMING WITH THE DRILL BIT WITH APPLIED PRESSURE. THE DRILL BIT ADVANCED FURTHER THAN IT WAS SUPPOSE TO. IT IS UNCERTAIN IF THE DRILL STOPPER WAS IN CORRECTLY BECAUSE THE DRILL STOPPER SLIPPED. THE DRILL BIT PERFORATED THE FEMORAL HEAD, THE SURGEON PULLED THE DRILL BIT BACK TO FINISH THE PROCEDURE SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRILL STOP DRILL STOP HTW SYNTHES BRANDYWINE 4777896

Patients

Seq Age Sex Outcome Treatment
1