OT VERIO IQ METER
Report
- Report Number
- 3008382007-2012-06463
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Report Date
- November 7, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. THE PRODUCT(S) HAVE NOT YET BEEN EVALUATED; LFS WILL EVALUATE IT/THEM AND INFORM FDA OF THE RESULTS IN A SUPPLEMENTAL REPORT.
FOLLOW-UP ((B)(4) 2012)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2012 RESPECTIVELY WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE ERROR COULD NOT BE REPRODUCED. THE TEST STRIPS WERE ALSO TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2012 THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) FROM UNITED STATES ALLEGING THAT THEIR ONETOUCH VERIO IQ METER WAS PROMPTING AN ERROR MESSAGE AND UNABLE TO GET A READING AFTER APPLYING BLOOD SAMPLE. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. REPLACEMENT PRODUCT WAS SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THEIR COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION(S): THE PATIENT CLAIMED THAT THE SUBJECT METER WAS PROMPTING AN ERROR MESSAGE AND UNABLE TO GET A READING AFTER APPLYING BLOOD SAMPLE. THERE IS NO EVIDENCE THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY SYMPTOMS, TREATMENT, OR BLOOD GLUCOSE VALUES SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3241896 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |