FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2841137 · Received November 21, 2012

Report

Report Number
3008382007-2012-06535
Event Type
Malfunction
Date Received
November 21, 2012
Report Date
November 6, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2012 THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) FROM THE UNITED STATES, ALLEGING THAT THEIR ONETOUCH VERIO IQ METER WAS PROMPTING SETTINGS MODE INSTEAD OF THE APPLY BLOOD SYMBOL WHEN TRYING TO TEST BLOOD GLUCOSE. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. BASED ON THE INFORMATION PROVIDED, THEIR COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION(S): THE PATIENT CLAIMED THAT THE SUBJECT METER WAS PROMPTING SETTINGS MODE INSTEAD OF THE APPLY BLOOD SYMBOL WHEN TRYING TO TEST BLOOD GLUCOSE. THERE IS NO EVIDENCE THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY SYMPTOMS, TREATMENT, OR BLOOD GLUCOSE VALUES SUGGESTIVE OF A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3354333

Patients

Seq Age Sex Outcome Treatment
1