SMALL CLIP APPLIER INSTRUMENT
Report
- Report Number
- 2955842-2012-01017
- Event Type
- Malfunction
- Date Received
- November 20, 2012
- Date of Event
- October 22, 2012
- Report Date
- October 22, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE CLIP APPLIER GRIP IS BROKEN AND WAS RETURNED WITH INSTRUMENT. THE BROKEN PIECE MEASURED APPROXIMATELY ABOUT 0.050 X 0.138. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO THE APPLICATION OF FORCE WHILE TRYING TO INSTALL A CLIP. THE BROKE ANGLE CAN BE REPLICATED ON AN IN-HOUSE TEST INSTRUMENT OF THE SAME TYPE WHEN A APPLYING A LOAD ON THE FIXTURE THAT CONTAINS THE CLIPS DURING INSTALLATION OF A CLIP ONTO THE GRIPS. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING SET-UP OF A DA VINCI S SURGICAL PROCEDURE ONE OF THE TIPS ON THE SMALL CLIP APPLIER INSTRUMENT BROKE OFF AND SEPARATED FROM THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SMALL CLIP APPLIER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420003-06 | S10111003 033 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |