HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2012-05693
- Event Type
- Malfunction
- Date Received
- November 20, 2012
- Date of Event
- November 3, 2012
- Report Date
- November 4, 2012
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THIS COMPLAINT FOR A REPORT OF A SYSTEM ERROR 2240 WAS CONFIRMED. THE ROOT CAUSE WAS USE ERROR - OPEN CLAMP. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.
THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN TUBING) AND SYSTEM ERROR 2367 WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 3 OF 5. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE PATIENT LINE HAD BEEN PROPERLY PRIMED PRIOR TO CONNECTING AND NO PATIENT EXTENSIONS WERE IN USE. THE PATIENT HAD NOT DISCONNECTED PRIOR TO THE ALARM. ALL OF THE BAGS WERE PROPERLY CONNECTED. THE SUPPLIES HAD NOT BEEN DAMAGED BY AN OUTLET PORT CLAMP OR ASSIST DEVICE. THERE WAS AN OPEN CLAMP ON AN UNUSED LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE TO THE REGISTERED NURSE (RN) AND ADVISED TO USE MANUAL BAGS TOP COMPLETE THERAPY. THE RN STATED THAT SHE DID NOT HAVE ANY TO USE. THE TSR ADVISED THE RN TO LET THE PATIENT'S PERITONEAL DIALYSIS RN KNOW ABOUT THE ISSUE. PROPER PROCEDURES WERE REVIEWED, AND THERE WERE NO SAMPLES AVAILABLE. THE LOT NUMBER WAS NOT AVAILABLE. THE HC WAS OKAY. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 92 YR | HOMECHOICE |