FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2840790 · Received November 20, 2012

Report

Report Number
2531779-2012-13813
Event Type
Malfunction
Date Received
November 20, 2012
Report Date
October 25, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1. (B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE PUMP WAS PROGRAMMED ON A ONE UNIT PER HOUR BASAL RATE AND EXERCISED FOR 24 HOURS AND NO UNPROGRAMMED BOLUSES WERE DELIVERED OR RECORDED IN THE PUMP HISTORY DURING THIS TIME. NO HYPERSENSITIVE KEYS WERE OBSERVED ON KEYPAD. A NORMAL 10 UNIT BOLUS AND A 10 UNIT AUDIO BOLUS WERE PERFORMED SUCCESSFULLY AND BOTH WERE ACCURATELY RECORDED IN THE PUMP HISTORY. THE KEYPAD WAS TESTED AND ALL KEYS ARE RESPONSIVE TO USER INPUT. THE COMPLAINT WAS NOT ABLE TO BE DUPLICATED DURING THE INVESTIGATION PROCESS.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

ON (B)(6) 2012 THE REPORTER CONTACTED ANIMAS TO REPORT THE PUMP HISTORY LISTED BOLUSES AND TIMES THAT HAD NOT BEEN PROGRAMMED INTO THE PUMP BY THE USER, AND SHE NOTED THESE BOLUSES HAD NOT BEEN DELIVERED. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR