FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2840764 · Received November 20, 2012

Report

Report Number
1416980-2012-05679
Event Type
Malfunction
Date Received
November 20, 2012
Date of Event
October 31, 2012
Report Date
October 31, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THAT THE YELLOW COIL CAP HAD SEPARATED FROM THE COVER. THE ROOT CAUSE WAS DETERMINED TO BE A MANUFACTURING DEFECT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD A COIL CAP SEPARATE FROM THE HOUSING. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12C051

Patients

Seq Age Sex Outcome Treatment