FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME

MDR report key: 2840174 · Received November 13, 2012

Report

Report Number
1037905-2012-00633
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 16, 2012
Report Date
October 17, 2012
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
KNS
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVAL: OUR EVAL OF THE RETURNED DEVICE WAS ABLE TO CONFIRM THE REPORT OF INCORRECT CUTTING WIRE ORIENTATION. DURING THE VISUAL ANALYSIS IT WAS NOTED THAT THE CUTTING WIRE WAS INTACT AND DID NOT EXHIBIT SIGNS OF DEFORMATION. THE TIP OF THE SPHINCTEROTOME WAS DEFORMED IN SHAPE PRESENTING A SLIGHTLY TWISTED APPEARANCE. DURING THE LAB ANALYSIS, THE SPHINCTEROTOME WAS ADVANCED THROUGH A DUODENOSCOPE THAT IS PLACED IN A SIMULATED BILIARY POSITION. THE DUODENOSCOPE HAS AN ACCESSORY CHANNEL THAT IS 4.2 MM IN DIAMETER (MODEL NUMBER OLYMPUS TJF-160V). THE CATHETER EXITED THE ENDOSCOPE WITH THE CUTTING WIRE FACING 1:30 O'CLOCK - ACCESS TO THE PAPILLA WAS ACHIEVABLE. THE DEVICE WAS THEN BOWED AND THE CUTTING WIRE WAS FACING 2:00 O'CLOCK . (APPROPRIATE ORIENTATION IS (B)(4)). IT WAS NOTED THAT ENTRY INTO THE SIMULATED PAPILLA WAS DIFFICULT GIVING AN INITIAL INDICATION OF MIS-ORIENTATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LAB SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LAB ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO THE DEVICE." OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULT AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING THE PROCEDURE, A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME WAS USED. THERE WAS A LEFT LATERAL DEVIATION WITH THIS PRODUCT [INCORRECT CUTTING WIRE ORIENTATION]. ANOTHER PRODUCT WAS USED TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME KNS, ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES KNS WILSON-COOK MEDICAL INC W3181663

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS TJF ENDOSCOPE (UNK MODEL NUMBER)