FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2840104 · Received November 20, 2012

Report

Report Number
1416980-2012-05655
Event Type
Malfunction
Date Received
November 20, 2012
Date of Event
October 28, 2012
Report Date
October 29, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED PROBLEM OF AN "INSERT SLIDE CLAMP" ALARM WAS CONFIRMED DURING EVALUATION. THE CAUSE OF THE PROBLEM WAS LOOSE SAFETY CLAMP SENSOR CONTACTS. THE SAFETY CLAMP SENSOR CONTACTS WERE CLEANED AND RE-SOLDERED TO FIX THE PROBLEM.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE/FLOGARD INFUSION PUMP WITH AN INSERT SIDE CLAMP ALARM. THE EVENT WAS REPORTED TO HAVE OCCURRED UPON POWER UP IN THE EMERGENCY ROOM. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.