FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 2840008 · Received November 20, 2012

Report

Report Number
2031642-2012-00508
Event Type
Malfunction
Date Received
November 20, 2012
Report Date
November 12, 2012
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

FACTORY ANALYSIS OF A RETURNED POWER MANAGEMENT PCB BOARD REVEALED FAILED COMPONENTS ON THE BOARD RESULTING IN A 35V SUPPLY FAILURE. THE NOTED FAILURE CAN RESULT IN A SHUTDOWN OF THE VENTILATOR DURING USE IN NORMAL VENTILATION MODE. UPON LOSS OF POWER, A POWER FAIL ALARM WILL ACTIVATE AND ALTERNATIVE VENTILATION SHOULD BE PROVIDED.