ERBE APC
Report
- Report Number
- 9610614-2012-00018
- Event Type
- Injury
- Date Received
- November 9, 2012
- Report Date
- November 9, 2012
- Manufacturer
- ERBE ELEKTROMEDIZIN GMBH
- Product Code
- GEI
- PMA / PMN Number
- K963189
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
AS REPORTED, IT APPEARS THAT COMBUSTIBLE GASSES (E.G., METHANE AND/OR HYDROGEN) WERE AT SUCH A CONCENTRATION THAT WHEN DIATHERMY WAS APPLIED AN EXPLOSION OCCURRED. ALTHOUGH VERY RARE, THE COMPLICATION CAN OCCUR. THEREFORE, THERE IS A WARNING IN THE USER MANUAL INVOLVING THE COMPLICATION AND INSTRUCTION THAT ENDOGENOUS GASES MUST BE REMOVED. NO TRENDS HAVE BEEN IDENTIFIED WITH THIS INCIDENT. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS REPORT.
A PT INCIDENT WAS REPORTED IN A CLINICAL ABSTRACT, "EXPLOSION FROM ARGON CAUTERY DURING PROCTOILEOSCOPY OF A PT WITH A COLECTOMY" (CLINICAL GASTROENTEROLOGY AND HEPATOLOGY 2012;10:1176-1178) INVOLVING ERBE EQUIPMENT [I.E., IN ALL PROBABILITY AN ERBE SYSTEM; ARGON PLASMA COAGULATOR (APC) MODEL APC 300 WITH AN ELECTROSURGICAL UNIT (ESU/GENERATOR) MODEL ICC 200 E/A, PART NUMBER 10128-205; SERIAL NUMBERS UNK]. THE EQUIPMENT SETTINGS WERE APC FORCED MODE (PRESUMABLY), 60 WATTS, 0.8 LITERS/MINUTE GAS FLOW. THE ABSTRACT SUMMARY WAS AS FOLLOWS: "WE REPORT A UNIQUE CASE OF A (B)(6) WOMAN WITH GARDNER'S SYNDROME WHO HAD A SUBTOTAL COLECTOMY WITH ILEOPROCTOSTOMY. SINCE THEN, SHE HAS UNDERGONE 12 UNCOMPLICATED PROCTOILESCOPIES, EACH TIME WITH ARGON PLASMA COAGULATION ABLATION OF SMALL POLYPS WITHOUT ANY BOWEL PREPARATION. HOWEVER, DURING THE MOST RECENT PROCEDURE, WHEN WE ATTEMPTED TO CAUTERIZE SOME RECTAL POLYPS, AN IMMEDIATE EXPLOSION OCCURRED, LEADING TO MULTIPLE RECTAL AND ILEAL PERFORATIONS THAT REQUIRED SURGICAL REPAIR WITH A TEMPORARY END ILEOSTOMY. THIS EVENT SUGGESTS THAT BACTERIAL FERMENTATION OF COLONIC CONTENT OR VISIBLE FECES IS NOT NECESSARY FOR COMBUSTION BECAUSE WE OBSERVED A CAUTERY-RELATED EXPLOSION IN THE ABSENCE OF A COLON. THIS CASE SHOWS THE NEED FOR ADEQUATE BOWEL PREPARATION IF CAUTERY IS TO BE USED, EVEN IN PTS WHO HAVE UNDERGONE A COLECTOMY."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ERBE APC | ARGON PLASMA COAGULATOR | GEI | ERBE ELEKTROMEDIZIN GMBH | APC 300 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Hospitalization| L| R |