FDA Adverse Event Injury Summary report: N

BBL THIOGLYCOLLATE MEDIUM, ENRICHED WITH VITA

MDR report key: 2836892 · Received October 31, 2012

Report

Report Number
1119779-2012-00016
Event Type
Injury
Date Received
October 31, 2012
Date of Event
October 1, 2012
Report Date
October 1, 2012
Manufacturer
BD DIAGNOSTIC SYSTEMS
Product Code
JSG
PMA / PMN Number
K803023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER HAS EVALUATED THE STAFF'S SPECIMEN PROCESSING PROCEDURE AND SWABBED VARIOUS AREAS OF THE LAB, STAFF AND OPERATING ROOM. HE STATES THAT THE STAFF'S PROCESSING WAS WITHIN PROTOCOL AND THE CONTAMINATE DID NOT GROW FROM THE ENVIRONMENTAL SAMPLES THAT HE TOOK. THE CUSTOMER INCUBATED UNINOCULATED VIALS AND FOUND NO CONTAMINATION EVIDENT. THE CUSTOMER ALSO STATED THAT THEY ARE REAUTOCLAVING THE MEDIA AND HAVE NOT SEEN FURTHER CONTAMINATION. CUSTOMER WAS MADE AWARE AT THIS TIME THAT THE MEDIA IS NOT VALIDATED FOR ADDITIONAL AUTOCLAVE CYCLES AND MAY FAIL TO FUNCTION AS INTENDED. CUSTOMER ALSO RULED OUT THEIR STERILE PIPETTES AND SWABS AS POSSIBLE SOURCES. THE BATCH HISTORY RECORD WAS REVIEWED AND ALL RESULTS ARE SATISFACTORY AND WITHIN NORMAL RANGES. NO RETURNS WERE SENT BY THE CUSTOMER. RETENTION SAMPLES WERE GRAM STAINED, WET PREPPED, INCUBATED AT 20-25C AND 30-35C THEN SUBCULTURED TO SABOURAD DEXTROSE AGAR. NO ORGANISMS WERE NOTED AND NO GROWTH WAS OBSERVED TO DATE. NO TRENDS HAVE BEEN NOTED TO DATE IN THE LOTS REPORTED BY THE CUSTOMER. MDRS WERE GENERATED ON EACH LOT OF MEDIA REPORTED BY THIS SINGLE CUSTOMER AS THE CUSTOMER WAS UNABLE TO DISTINGUISH WHICH LOT WAS USED WHEN A POTENTIAL FALSE RESULT WAS REPORTED.

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT THAT THEY HAVE BEEN HAVING AN ONGOING CONTAMINATION EVENT WITH TWO LIQUID PRODUCTS FROM BD. THE CUSTOMER IS SEEING CONTAMINATION WITH CANDIDA PARAPSILOSIS. THEY HAVE BEEN HAVING THIS ISSUE FOR SEVERAL MONTHS WITHOUT BEING ABLE TO TRACK THE SOURCE BUT DID NOT PROVIDE SPECIFIC DATING. THE COMMON ELEMENT IS THE USE OF BD THIOGLYCOLLATE BROTH AND BRAIN HEART INFUSION BROTH. NO CONTAMINATION WAS SEEN ON THE GRAM STAIN OR ON THE ORIGINAL MEDIA PLATES. THE CONTAMINATION WAS SEEN AFTER THE INOCULATION OF THE TUBES. CUSTOMER'S PROCESS IS TO INOCULATE THE BHI, THEN INOCULATE THE BHI SUSPENSION INTO THE THIOGLYCOLLATE. THE CUSTOMER DOES NOT KNOW THE NUMBER OF PATIENTS THAT WERE AFFECTED OR SPECIFIC LOT NUMBERS OF PRODUCT, BUT STATES THAT (B)(4) PATIENTS IN TOTAL WERE AFFECTED THROUGHOUT ALL OF THE LOTS. THE AFFECTED SPECIMENS WERE PRIMARILY FROM STERILE SITES BUT THE CUSTOMER COULD NOT PROVIDE FURTHER DETAILS. THE SPECIMENS WERE COLLECTED AT THE SAME FACILITY IN THE OPERATING ROOM OR EMERGENCY ROOM. THE AFFECTED PATIENTS WERE INITIALLY TREATED WITH ANTIFUNGAL MEDICATION. NO FURTHER PATIENT INFORMATION IS AVAILABLE. THESE ISSUES WILL BE DOCUMENTED ON MDRS 1119779-2012-00013 TO 1119779-2012-00019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BBL THIOGLYCOLLATE MEDIUM, ENRICHED WITH VITA JSG BD DIAGNOSTIC SYSTEMS 2115010

Patients

Seq Age Sex Outcome Treatment
1 Other