FENESTRATED BIPOLAR FORCEPS
Report
- Report Number
- 2955842-2012-00966
- Event Type
- Other
- Date Received
- November 16, 2012
- Date of Event
- October 17, 2012
- Report Date
- October 19, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- NURSE
Narratives
ISI CONTACTED INITIAL REPORTER, REGISTER NURSE (B)(6). BASED ON THE INFORMATION PROVIDED BY (B)(6), DUAL CONSOLES WERE USED FOR THIS PROCEDURE, SURGEON DR (B)(6) WAS USING ONE CONSOLE AND THE RESIDENT SURGEON WAS USING THE SECOND CONSOLE. SHE INDICATED THAT THERE WAS A COLLISION BETWEEN THE FENESTRATED BIPOLAR INSTRUMENT USED BY DR (B)(6) AND THE PROGRASP INSTRUMENT USED BY THE RESIDENT SURGEON. AS THE SURGEONS ATTEMPTED TO SEPARATE BOTH INSTRUMENTS, THEY OBSERVED THAT THE FENESTRATED BIPOLAR FORCEPS JAW BROKE. IT TOOK APPROXIMATELY ONE HOUR TO LOCATE AND REMOVE THE BROKEN PIECE FROM THE PATIENT. AN X-RAY HAD TO BE PERFORMED TO LOCATE THE INSTRUMENT PIECE; THE PIECE WAS LOCATED IN THE PATIENT'S PELVIS. PATIENT DID NOT SUFFER ANY ADDITIONAL COMPLICATIONS FROM THIS INCIDENT OR FROM THE EXPOSURE TO ADDITIONAL ANESTHESIA. PATIENT WAS RELEASED FROM THE HOSPITAL AS PLANNED. THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS RETURNED WITH ONE GRIP JAW BROKEN AT DISTAL CLEVIS GRIP BASE. A PIECE OF APPROXIMATELY 0.185 X 0.209 HAD BROKEN OFF FROM THE INSTRUMENT. THERE WAS NO DAMAGE FOUND ON THE CABLE OR THE WIRE. INSTRUMENT PASSED ELECTRICAL CONTINUITY TEST. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE BIPOLAR FORCEPS INSTRUMENT BROKE AND HALF OF THE JAW FELL INTO THE PATIENT. THE SURGEON WAS ABLE TO LOCATE THE BROKEN JAW PIECE IN THE PELVIS, THE PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FENESTRATED BIPOLAR FORCEPS | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10120813 710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | DA VINCI SI SYSTEM INSTRUMENTS| ACCESSORIES |