FDA Adverse Event Injury Summary report: N

AMS GREENLIGHT HPS FIBER

MDR report key: 2835758 · Received October 22, 2012

Report

Report Number
2937094-2012-01100
Event Type
Injury
Date Received
October 22, 2012
Date of Event
June 29, 2012
Report Date
July 11, 2012
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE MANUFACTURER ASSIGNED MODEL NUMBER 0010-2093 IS DESIGNATED FOR (B)(4) DISTRIBUTION. THIS REPORT IS BEING SUBMITTED AS THE MODEL 0010-2093 IS IDENTICAL I DESIGN AND MANUFACTURE TO THE COMMERCIALLY AVAILABLE MODEL 0010-2090 ANGLED DELIVERY DEVICE, GREENLIGHT (K062719). FIBER ANALYSIS: THE FIBER CAP REMAINS INTACT AND ATTACHED AND DRILLED THROUGH; EXHIBITS DETRITUS, CHAR, DEVITRIFICATION, AND MELTED GLASS. CAP CONDITION WOULD RESULT IN REDUCED VAPORIZATION EFFICIENCY. THIS MAY CAUSE FORWARD FIRING. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SIDE-FIRING FIBER WAS NOTICED TO HAVE COMMENCED FORWARD FIRING AT 111,991 JOULES DURING A PROSTATE PROCEDURE. THE PROCEDURE WAS COMPLETED BY TURP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AMS GREENLIGHT HPS FIBER POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS 0010-2093 148J

Patients

Seq Age Sex Outcome Treatment
1 Other GREENLIGHT HPS LASER SYSTEM AND ACCESSORIES