AMS GREENLIGHT HPS FIBER
Report
- Report Number
- 2937094-2012-01104
- Event Type
- Injury
- Date Received
- October 22, 2012
- Date of Event
- May 9, 2012
- Report Date
- June 1, 2012
- Manufacturer
- AMERICAN MEDICAL SYSTEMS
- Product Code
- GEX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE MANUFACTURER ASSIGNED MODEL NUMBER 0010-2093 IS DESIGNATED FOR (B)(4) DISTRIBUTION. THIS REPORT IS BEING SUBMITTED AS THE MODEL 0010-2093 IS IDENTICAL IN DESIGN AND MANUFACTURE TO THE COMMERCIALLY AVAILABLE MODEL 0010-2090 ANGLED DELIVERY DEVICE, GREENLIGHT (K062719). FIBER ANALYSIS: THE FIBER CAP REMAINS INTACT AND ATTACHED AND DRILLED THROUGH; EXHIBITS DETRITUS, CHAR, DEVITRIFICATION, AND MELTED GLASS. CAP CONDITION WOULD RESULT IN REDUCED VAPORIZATION EFFICIENCY. THIS MAY CAUSE FORWARD FIRING. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER. REFERENCE: 2937094-2012-01103.
IT WAS REPORTED THAT THE FIRST FIBER TIP WAS DAMAGED AT 31,614; SECOND FIBER TIP WAS DAMAGED AT 87,614 JOULES DURING A PROSTATE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH TURP. NO REPORT OF PATIENT INJURY. THIS REPORT IS FOR THE SECOND FIBER AND TURP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMS GREENLIGHT HPS FIBER | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS | 0010-2093 | 148J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | GREENLIGHT HPS LASER SYSTEM AND ACCESSORIES |