FDA Adverse Event
Other
Summary report: N
MULTISCAN TABLE
MDR report key: 2835359
·
Received November 5, 2012
Report
- Report Number
- 1932056-2012-00006
- Event Type
- Other
- Date Received
- November 5, 2012
- Date of Event
- October 25, 2012
- Report Date
- November 5, 2012
- Manufacturer
- MEDICAL POSITIONING, INC.
- Product Code
- LGX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
THE BED WAS IN IT'S LOWEST HEIGHT AND IN A HORIZONTAL POSITION. AS THE TECHNICIAN BEGAN TO RAISE THE HEIGHT OF THE BED, THE HEAD PORTION OF THE BED TILTED DOWN DUE TO THE TRENDELENBURG ACTUATOR CLEVIS BREAKING AND RELEASING. AS A RESULT, THE PATIENT SLID TOWARD THE HEAD OF THE BED FORCING HIS LEGS UPWARD BUT DID NOT FALL OFF THE BED. THE PATIENT SUFFERS WITH BACK PAIN WHICH SEEM TO BE AGGRAVATED BY THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MULTISCAN TABLE | MULTISCAN TABLE | LGX | MEDICAL POSITIONING, INC. | 8083 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Other |