FDA Adverse Event Other Summary report: N

MULTISCAN TABLE

MDR report key: 2835359 · Received November 5, 2012

Report

Report Number
1932056-2012-00006
Event Type
Other
Date Received
November 5, 2012
Date of Event
October 25, 2012
Report Date
November 5, 2012
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
LGX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

THE BED WAS IN IT'S LOWEST HEIGHT AND IN A HORIZONTAL POSITION. AS THE TECHNICIAN BEGAN TO RAISE THE HEIGHT OF THE BED, THE HEAD PORTION OF THE BED TILTED DOWN DUE TO THE TRENDELENBURG ACTUATOR CLEVIS BREAKING AND RELEASING. AS A RESULT, THE PATIENT SLID TOWARD THE HEAD OF THE BED FORCING HIS LEGS UPWARD BUT DID NOT FALL OFF THE BED. THE PATIENT SUFFERS WITH BACK PAIN WHICH SEEM TO BE AGGRAVATED BY THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MULTISCAN TABLE MULTISCAN TABLE LGX MEDICAL POSITIONING, INC. 8083

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other