FDA Adverse Event Malfunction Summary report: N

FREEDOM EVO-2 100

MDR report key: 2834953 · Received November 9, 2012

Report

Report Number
3003402518-2012-00001
Event Type
Malfunction
Date Received
November 9, 2012
Date of Event
October 14, 2012
Report Date
November 8, 2012
Manufacturer
TECAN SCHWEIZ AG
Product Code
JQW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

TECAN EVALUATED THE LIQUID HANDLING ARM NOTING THAT THE TANTAL CONDENSER CAPACITOR C41 OVERHEATED WHICH RESULTED IN SMOKE DEVELOPMENT AND DESTRUCTION OF PCB BOARDS. THE ASSEMBLY HAS BEEN SENT TO SUPPLIER FOR FURTHER INVESTIGATION. THE RESULTS FROM THAT INVESTIGATION WILL BE DOCUMENTED IN A FOLLOW UP REPORT. THIS APPEAR TO BE AN ISOLATED EVENT. THERE WAS NO FACILITY DAMAGE OR INJURY TO ANY LABORATORY PERSONNEL.

Description of Event or Problem · 1

TECAN WAS NOTIFIED BY DISTRIBUTOR IN (B)(6) OF AN ERROR WHERE THE LIQUID HANDLING ARM ON THE EVO 100 COLLIDED WITH THE WASH STATION DURING MOVEMENT ALONG THE X AXIS. OPERATOR REPORTED A SPARK FROM THE BACK OF THE LIQUID HANDLING ARM AND INSTRUMENT WAS SHUT DOWN. THERE WAS NO INJURY TO ANY PERSONNEL OR DAMAGE TO THE FACILITY. EVO INSTRUMENT WAS NO LONGER OPERATIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREEDOM EVO-2 100 PIPETTING STATION FOR CLINICAL USE JQW TECAN SCHWEIZ AG EVO-2 100

Patients

Seq Age Sex Outcome Treatment
1 Other