FDA Adverse Event Malfunction Summary report: N

J-VAC BULB RESERVOIR

MDR report key: 2834614 · Received November 15, 2012

Report

Report Number
2210968-2012-07366
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
October 25, 2012
Report Date
October 25, 2012
Manufacturer
ETHICON, INC.
Product Code
KOG
PMA / PMN Number
K953655
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ANTI-REFLUX VALVE DETACHMENT. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. ADDITIONAL INFORMATION: THE ACTUAL DEVICE BATCH NUMBER ASSOCIATED WITH THIS EVENT IS NOT KNOWN. THE INTERNATIONAL AFFILIATE REPORTS THE FOLLOWING POSSIBLE BATCH NUMBERS: BATCH J1227917; BATCH J1227920.

Additional Manufacturer Narrative · 1

CONCLUSION: THE ACTUAL PRODUCT WAS RECEIVED FOR EVALUATION. THE VALVE IS DETACHED AND INSIDE THE RESERVOIR. THE VALVE'S SLIT IS CLOSED. INFORMATION: THE ACTUAL DEVICE BATCH NUMBER ASSOCIATED WITH THIS EVENT IS NOT KNOWN. THE INTERNATIONAL AFFILIATE REPORTS THE FOLLOWING POSSIBLE BATCH NUMBERS: BATCH # J1227917; MFG DATE 2012-03; EXPIRATION DATE 2017-03. BATCH J1227920; MFG DATE 2012-04; EXPIRATION DATE 2017-04. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS FOR THE POSSIBLE BATCH NUMBERS WAS CONDUCTED AND THE BATCHES MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT PROCEDURE ON (B)(6) 2012 AND A RESERVOIR WAS CONNECTED TO THE DRAIN AT THE PERICARDIUM. ON (B)(6) 2012, THE STAFF FOUND THAT THE ANTI-REFLUX VALVE HAD DETACHED AND FELL INTO THE RESERVOIR. ANOTHER LIKE DEVICE WAS USED WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 J-VAC BULB RESERVOIR WOUND DRAINAGE SYSTEM KOG ETHICON, INC. NA NI

Patients

Seq Age Sex Outcome Treatment
1