J-VAC BULB RESERVOIR
Report
- Report Number
- 2210968-2012-07366
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- October 25, 2012
- Report Date
- October 25, 2012
- Manufacturer
- ETHICON, INC.
- Product Code
- KOG
- PMA / PMN Number
- K953655
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ANTI-REFLUX VALVE DETACHMENT. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. ADDITIONAL INFORMATION: THE ACTUAL DEVICE BATCH NUMBER ASSOCIATED WITH THIS EVENT IS NOT KNOWN. THE INTERNATIONAL AFFILIATE REPORTS THE FOLLOWING POSSIBLE BATCH NUMBERS: BATCH J1227917; BATCH J1227920.
CONCLUSION: THE ACTUAL PRODUCT WAS RECEIVED FOR EVALUATION. THE VALVE IS DETACHED AND INSIDE THE RESERVOIR. THE VALVE'S SLIT IS CLOSED. INFORMATION: THE ACTUAL DEVICE BATCH NUMBER ASSOCIATED WITH THIS EVENT IS NOT KNOWN. THE INTERNATIONAL AFFILIATE REPORTS THE FOLLOWING POSSIBLE BATCH NUMBERS: BATCH # J1227917; MFG DATE 2012-03; EXPIRATION DATE 2017-03. BATCH J1227920; MFG DATE 2012-04; EXPIRATION DATE 2017-04. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS FOR THE POSSIBLE BATCH NUMBERS WAS CONDUCTED AND THE BATCHES MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT PROCEDURE ON (B)(6) 2012 AND A RESERVOIR WAS CONNECTED TO THE DRAIN AT THE PERICARDIUM. ON (B)(6) 2012, THE STAFF FOUND THAT THE ANTI-REFLUX VALVE HAD DETACHED AND FELL INTO THE RESERVOIR. ANOTHER LIKE DEVICE WAS USED WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | J-VAC BULB RESERVOIR | WOUND DRAINAGE SYSTEM | KOG | ETHICON, INC. | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |