RESERVOIR 3ML
Report
- Report Number
- 3004209178-2012-90792
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- November 5, 2012
- Report Date
- November 5, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
INSPECTED THREE SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATIONS. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATIONS. NO LEAKAGE ANOMALY OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WAS FOUND. RESERVOIR CONNECTED AND LOCKED IN PLACE PROPERLY.
THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR WAS LEAKING PAST THE O-RINGS DURING NORMAL USE. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESERVOIR 3ML | FRN | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A | H8187919 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |