PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2012-90777
- Event Type
- Injury
- Date Received
- November 15, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 2, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OVER 500MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND BOLUS WIZARD SETTINGS WERE CORRECT. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND PASSED. INSTRUCTED THE CUSTOMER TO REMOVE THE CANNULA AND IT WAS NOT BENT. THE CUSTOMER STATED THAT THE HEALTH CARE PROVIDER REQUESTED THE DEVICE BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization |